RCF to RPM Calculator (Centrifuge g-Force Conversion)
Enter the rotor radius, then either the speed or the g-force, and get the other one. Useful when a protocol gives ×g but your centrifuge only lets you set rpm.
Inputs
Convert
r max: distance from the rotor axis to the tube bottom
Results
- Speed
- 11,583 rpm
Formula
RCF (×g) = 1.118 × 10⁻⁵ × r (cm) × rpm²
rpm = √(RCF ÷ (1.118 × 10⁻⁵ × r))
Tip: Use r max, the distance from the rotor axis to the tube bottom. Results are rounded to whole numbers so you can enter them directly.
You got this step right, but a whole experiment has dozens of linked parameters. Change one dilution factor and everything downstream has to be recalculated. In Vara the parameters of your whole experiment are connected, so changing one recalculates the rest automatically. Manage 3 experiments for free.
How the formula works
- Centrifugal force comes from centripetal acceleration: a = ω² × r, where ω is the angular velocity in rad/s and r is the distance from the axis of rotation.
- rpm is revolutions per minute. Convert it to angular velocity with ω = 2π × rpm ÷ 60.
- RCF says how many times stronger that acceleration is than gravity: RCF = ω² × r ÷ g, with g = 9.80665 m/s².
- With r entered in cm, divide by 100 to get meters. Combining the constants, (2π ÷ 60)² ÷ (100 × 9.80665) ≈ 1.118 × 10⁻⁵, which gives RCF = 1.118 × 10⁻⁵ × r (cm) × rpm².
- Solving for speed: rpm = √(RCF ÷ (1.118 × 10⁻⁵ × r)). Force goes with the square of the speed, so doubling the rpm gives four times the ×g.
- Which radius: different parts of the tube sit at different distances from the axis. The tube bottom is farthest out (r max), the liquid surface is closest (r min), and the midpoint between them is the average radius (r av). At a given rpm the force is highest at r max. Manufacturers and most protocols quote ×g at r max; some protocols, more often in ultracentrifugation, quote it at r av. Use whichever radius your protocol specifies.
- Rounding: this calculator rounds speed to the nearest whole rpm, and rounds force to a whole number at 10 ×g and above, or to one decimal place below 10 ×g.
Worked examples
Microcentrifuge: what is 12,000 rpm at r = 8.4 cm in ×g?
- Plug in: RCF = 1.118 × 10⁻⁵ × 8.4 × 12000².
- 12000² = 1.44 × 10⁸.
- 1.118 × 10⁻⁵ × 8.4 × 1.44 × 10⁸ = 13,523.328.
- Rounded: 13,523 ×g. The 8.4 cm radius is only an example; use the r max printed on your rotor or in its manual.
Swing-bucket rotor: what rpm gives 500 ×g at r = 15 cm?
- Use the reverse formula: rpm = √(500 ÷ (1.118 × 10⁻⁵ × 15)).
- 1.118 × 10⁻⁵ × 15 = 1.677 × 10⁻⁴.
- 500 ÷ (1.677 × 10⁻⁴) ≈ 2,981,515.
- √2,981,515 ≈ 1726.7, which rounds to 1727 rpm.
- If your centrifuge only steps in 100 rpm, 1700 rpm gives about 485 ×g, slightly under target; if it steps in 10 rpm, 1730 rpm gives about 502 ×g. Whether that deviation is acceptable is up to your protocol.
Common mistakes
Using the wrong rotor radius
Plugging in r min, or the radius of a different rotor. At the same 12,000 rpm, r = 8.4 cm gives 13,523 ×g while 4.5 cm gives only about 7245 ×g, a little over half. Check whether your protocol means r max or r av, then look up that radius for the rotor you are using.
Entering ×g as rpm
The protocol says 5000 ×g and the centrifuge gets set to 5000 rpm. Many centrifuges can switch the display between rpm and RCF, so check which unit is showing before you set it.
Keeping the same rpm after changing rotors
Swap the rotor on the same centrifuge and the radius changes, so the same rpm no longer gives the same ×g. Record ×g in your notes, or the rotor model together with the rpm, and recalculate when you change rotors.
Treating fixed-angle and swing-bucket rotors as the same
In a fixed-angle rotor the tube is held at an angle and the pellet forms on the outer wall near the bottom, with a short sedimentation path. In a swing-bucket rotor the tube swings out horizontal, the pellet forms at the center of the tube bottom, and r max is usually larger. At the same ×g, switching rotor types can change where the pellet sits and how long the spin needs, so if your protocol names a rotor type, follow it.
FAQ
My protocol says 12,000 ×g but my centrifuge only takes rpm. What do I set?+
How do I find the rotor radius?+
Why does the same rpm give a different g-force on another centrifuge?+
Roughly how many ×g is 3000 rpm?+
Do I need to change the spin time after converting?+
What is the formula to convert RCF to rpm?+
Why do protocols give ×g instead of rpm?+
Where does the 1.118 × 10⁻⁵ constant come from?+
How many ×g is 13,000 rpm?+
Related protocol guides
Vara Toolbox
- Plate layout designer
- Antibody Dilution Calculator (1:N)
- Solution Preparation Calculator (% w/v)
- PCR / qPCR Reaction Mix Calculator (Master Mix)
- Molarity Calculator (Mass, Molecular Weight, Volume, Concentration)
- Cell Seeding Calculator (Cells per Well and Seeding Density)
- Serial Dilution Calculator (Volumes and Concentration per Tube)
- Percent Solution Calculator (% w/v, % v/v, % w/w)